Model Number SMMA11R
Medical Device Problem Codes Positioning Failure (1158); Difficult to Insert (1316); Difficult to Remove (1528)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to bd for evaluation.Photos were returned for review.The company is still investigating the issue at this time.The device is labeled for single use.(b)(4).
 
Event or Problem Description
This report summarizes one malfunction.A review of the reported information indicated that model smma11r breast tissue marker allegedly experienced failure to insert / remove, failure to deploy, and device detachment.This information was received from one source.This malfunction did involve a patient with no reported patient injury.The patient was male and the age and weight were not reported.
 
Event or Problem Description
This report summarizes one malfunction.A review of the reported information indicated that model smma11r breast tissue marker allegedly experienced failure to insert / remove, failure to deploy, and device detachment.This information was received from one source.This malfunction did involve a patient with no reported patient injury.The patient was male and the age and weight were not reported.
 
Additional Manufacturer Narrative
H10: the lot number for the malfunction was provided and a lot history review was performed.The device has been returned for evaluation, as well as, a photo of the reported event.The investigation is confirmed for failure to deploy, difficult to remove, and detachment of device component, but is inconclusive for difficulty to insert.Based upon the available information, a definitive root cause is unknown.The device is labeled for single use.H10: g4, h6(2907- detachment of device or device component).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENOMARK ULTRA MARKER
Common Device Name
BREAST TISSUE MARKER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key9200135
Report Number2020394-2019-04042
Device Sequence Number11539898
Product Code NEU
UDI-Device Identifier00801741032073
UDI-Public(01)00801741032073
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model NumberSMMA11R
Device Catalogue NumberSMMA11R
Device Lot NumberVTCP00083
Initial Date Received by Manufacturer 09/30/2019
Supplement Date Received by Manufacturer03/31/2020
Initial Report FDA Received Date10/16/2019
Supplement Report FDA Received Date04/03/2020
Patient Sequence Number1
-
-