| Model Number |
C2TR01 |
| Medical Device Problem Code |
Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/10/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: 419688 lead, implanted: (b)(6) 2010; c2tr01 ipg, implanted: (b)(6) 2015.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the right ventricular (rv) lead had a gradual rise in impedance measurement causing a ¿drag¿ on the cardiac resynchronization therapy pacemaker (crt-p) battery.It was noted that the rv lead had possible lead mineralization and scarring at the rv ring.It was also noted that the left ventricular (lv) lead had high threshold measurements.The device, rv lead and lv lead remain in use.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information received noted reprogramming was done.
|
| |
|