Model Number C2TR01
Medical Device Problem Code Battery Problem (2885)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/10/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Concomitant medical products: 419688 lead, implanted: (b)(6) 2010; c2tr01 ipg, implanted: (b)(6) 2015.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the right ventricular (rv) lead had a gradual rise in impedance measurement causing a ¿drag¿ on the cardiac resynchronization therapy pacemaker (crt-p) battery.It was noted that the rv lead had possible lead mineralization and scarring at the rv ring.It was also noted that the left ventricular (lv) lead had high threshold measurements.The device, rv lead and lv lead remain in use.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information received noted reprogramming was done.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCRA CRT-P
Common Device Name
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH  1131
MDR Report Key9200534
Report Number9614453-2019-03543
Device Sequence Number3058201
Product Code NVZ
UDI-Device Identifier00613994753304
UDI-Public00613994753304
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P010015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/14/2017
Device Model NumberC2TR01
Device Catalogue NumberC2TR01
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/25/2019
Supplement Date Received by Manufacturer10/22/2019
Initial Report FDA Received Date10/16/2019
Supplement Report FDA Received Date10/23/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age82 YR
-
-