| Model Number |
FGS-0634 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
09/26/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the recorder failed to power on.A repeat procedure was done on the same day due to device malfunction with need for additional anesthesia.There was no reported patient outcome.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: this report is based on information provided by medtronic investigation personnel.The bravo recorder was received for evaluation.The returned sample did not meet specification as received by medtronic.The visual inspection found no issues.The customer reported the recorder failed to power on.The reported condition was confirmed.The investigation found that have a problem with turn on.A new firmware is resolved this problem.The investigation identified the cause of the reported event to be software bug.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|