MAUDE Adverse Event Report: COOK INC MODIFIED NOVY CORNUAL CANNULATION SET; MOV CATHETER, SALPINGOGRAPHY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC MODIFIED NOVY CORNUAL CANNULATION SET; MOV CATHETER, SALPINGOGRAPHY Back to Search Results
Model Number G17558
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/04/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported that during an irrigation and angiography on one side of the fallopian, a modified novy cornual cannulation set was used.After the distal tip of the introducing catheter was inserted into the patient it partially detached.The operator immediately removed the catheter from the patient, the partially detached tip was also removed.The procedure was then completed with another modified novy cornual cannulation set.No adverse effects to the patient were reported due to this occurrence and no portion of the device remained inside the patient.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unchanged, unknown or unavailable.Ec method code desc - 5: communication/interviews (4111).Investigation - evaluation: reviews of complaint history, device history record, manufacturing instructions, quality control data, and trend analysis were conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record found no non-conformances related to the reported failure mode.A review of complaint history records shows no other complaints associated with the complaint device lot.Because there are no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other related complaints from the lot have been received from the field, it was concluded that there is no evidence that nonconforming product exists in house or in field.The complainant did not return the complaint device to cook for investigation.A definitive cause of the reported event could not be determined.Per the quality engineering risk assessment, no further action is warranted.Cook medical will continue to monitor this device via the complaints database for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional patient or event information has been received since the last report was submitted on 17oct2019.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MODIFIED NOVY CORNUAL CANNULATION SET
Common Device Name
MOV CATHETER, SALPINGOGRAPHY
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key9204253
Report Number1820334-2019-02639
Device Sequence Number11488886
Product Code MOV
UDI-Device Identifier00827002175587
UDI-Public(01)00827002175587(17)210312(10)8659230
Combination Product (Y/N)N
Initial Reporter CountryCN
PMA/510(K) Number
K931476
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,other,use
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/12/2021
Device Model NumberG17558
Device Catalogue NumberJ-NCS-504070
Device Lot Number8659230
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/14/2019
Supplement Date Received by Manufacturer12/10/2019
Initial Report FDA Received Date10/17/2019
Supplement Report FDA Received Date12/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age32 YR
Patient Weight58
-
-