| Model Number |
ES28182400 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
09/18/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The lot number was reviewed for complaint trend, nonconforming reports and capa review.No trends were noted.Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional facts prompt us to alter or supplement any information or conclusions contained in the original mdr or in any prior supplemental reports, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
According to the available information.Functioning implant extruded through proximal corpora.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up was created to document the updated device information.A touch pump and two cylinders were received for evaluation.As examination of the device may not conclusively confirm or disprove the report of extrusion quality accepts the physician's observations as to the reason for surgical intervention.A review of the device history record confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database, nonconformances and capas revealed no trends for this lot.
|
| |
|