Model Number ES28182400
Device Problem Device Appears to Trigger Rejection (1524)
Patient Problem No Information (3190)
Event Date 09/18/2019
Event Type  Injury  
Manufacturer Narrative
The lot number was reviewed for complaint trend, nonconforming reports and capa review.No trends were noted.Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional facts prompt us to alter or supplement any information or conclusions contained in the original mdr or in any prior supplemental reports, a follow-up report will be submitted.
 
Event Description
According to the available information.Functioning implant extruded through proximal corpora.
 
Manufacturer Narrative
This follow-up was created to document the updated device information.A touch pump and two cylinders were received for evaluation.As examination of the device may not conclusively confirm or disprove the report of extrusion quality accepts the physician's observations as to the reason for surgical intervention.A review of the device history record confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database, nonconformances and capas revealed no trends for this lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TITAN TOUCH INFR ZERO ANG 18CM
Type of Device
INFLATABLE PENILE PROSTHESIS
Manufacturer (Section D)
COLOPLAST A/S
holtedam 1
humlebaek, 3050
DA  3050
MDR Report Key9204933
MDR Text Key162816636
Report Number2125050-2019-00882
Device Sequence Number1
Product Code FHW
UDI-Device Identifier05708932539074
UDI-Public05708932539074
Combination Product (y/n)N
PMA/PMN Number
P000006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 11/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberES28182400
Device Catalogue NumberES2818
Device Lot Number5537090
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/31/2019
Initial Date Manufacturer Received 09/18/2019
Initial Date FDA Received10/17/2019
Supplement Dates Manufacturer Received09/18/2019
Supplement Dates FDA Received11/20/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age72 YR
-
-