• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SP*2 F SIZG GD-POSTUP/SZ1.5-6; KNEE INSTRUMENT : SIZING/MEASURING INSTRUMENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SP*2 F SIZG GD-POSTUP/SZ1.5-6; KNEE INSTRUMENT : SIZING/MEASURING INSTRUMENTS Back to Search Results
Catalog Number 966131
Medical Device Problem Codes Mechanical Jam (2983); Naturally Worn (2988); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/26/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the holes for drill pins are frayed and doesn¿t allow a drill bit to pass through the hole without binding up.No surgical delay.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.(b)(4).Investigation summary: examination of the returned device confirmed the reported event of wear.The noted damage/wear is consistent with device use from normal use and servicing and the investigation did not establish a need for corrective action.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
KNEE INSTRUMENT : SIZING/MEASURING INSTRUMENTS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9205010
Report Number1818910-2019-110522
Device Sequence Number19590821
Product Code HWT
UDI-Device Identifier10603295246954
UDI-Public10603295246954
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number966131
Device Lot NumberH0305
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2019
Initial Date Received by Manufacturer 09/26/2019
Supplement Date Received by Manufacturer11/25/2019
Initial Report FDA Received Date10/17/2019
Supplement Report FDA Received Date11/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-