• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII PS INSERT SZ 3-4 9MM; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII PS INSERT SZ 3-4 9MM; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY Back to Search Results
Catalog Number 71420816
Medical Device Problem Code Connection Problem (2900)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/20/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a right knee replacement surgery, the 9 mm liner couldn't be implanted completely and was easy to fall off since there is a 1 mm gap between liner and tibia.Finally a 11 mm liner was used to complete the surgery.No injuries or delays reported.A delay greater than 30 min was reported.
 
Additional Manufacturer Narrative
The associated device was returned and evaluated.A visual inspection of the returned device could not confirm the stated failure mode.The device has signs of damage from the attempted insertion.The device was manufactured in 2019.A dimensional inspection was attempted but the device was too damaged from the attempted insertion to obtain accurate measurements.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII PS INSERT SZ 3-4 9MM
Common Device Name
PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9205347
Report Number1020279-2019-03673
Device Sequence Number10874531
Product Code JWH
UDI-Device Identifier03596010209573
UDI-Public03596010209573
Combination Product (Y/N)N
PMA/510(K) Number
K951987
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71420816
Device Lot Number19BT16043
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/25/2019
Initial Date Received by Manufacturer 09/20/2019
Supplement Date Received by Manufacturer09/20/2019
Initial Report FDA Received Date10/17/2019
Supplement Report FDA Received Date12/04/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age70 YR
-
-