| Model Number |
7134 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: used the first day of the month of the aware date as there was no date provided.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.A 3.00mm x 20mm emerge balloon catheter was selected for use.However, the balloon broke and had to be retrieved.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event: used the first day of the month of the aware date as there was no date provided.Device evaluated by mfr.: returned product consisted of an emerge balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was a complete separation at 23.6cm distal of the strain relief.There was contrast in the inflation lumen and balloon, and blood in the guidewire lumen.The balloon was loosely folded.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.A 3.00mm x 20mm emerge balloon catheter was selected fo use.However, the balloon broke and had to be retrieved.No patient complications were reported.
|
| |
|