| Medical Device Problem Code |
Image Display Error/Artifact (1304)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device has been returned and evaluated by product analysis on 02-jul-2019 with the following findings: the pump powered on and no lines were observed in the display.The complaint of lines in through the display was not duplicated during investigation.Unrelated to the complaint, the display was dim, faded and discolored.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.Report late due to transition from vmsr program.
|
| |
|
Event or Problem Description
|
|
On (b)(4) 2018, the reporter contacted animas, alleging a display (line through display) issue.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the issue may impact the user's ability to read some or all of the information on the screen which may result in over or under delivery.
|
| |
|