| Medical Device Problem Code |
Failure to Read Input Signal (1581)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has been returned and evaluated by product analysis on 15-jul-2019 with the following findings: no damage or peeling was observed to the keypad.The keypad was removed and no evidence of contamination was observed on button contacts.During testing, the pump failed to power on.The complaint of a keypad button response issue was not able to be adequately investigated due to no power to the pump.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.Report late due to transition from vmsr program.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, the reporter contacted animas, alleging a button/keypad (tactile changes/unresponsive) issue.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the issue has the ability to result in inadvertent or incorrect insulin delivery or the inability to use the pump.
|
| |
|