Catalog Number 136505100
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Edema (1820); Foreign Body Reaction (1868); Joint Dislocation (2374); Test Result (2695); No Code Available (3191)
Date of Event 05/16/2012
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Initial reporter occupation: lawyer.Patient code: no code available (3191) used to capture the surgical intervention, joint instability, and medical device removal.
 
Event or Problem Description
Ppf and medical records received.Ppf alleges dislocation and elevated metal ions.After review of medical records the patient was revised due recurrent dislocation.Operative notes reported, there was some metallosis as well as significant fluid anterior.There did not appear to be any obvious impingement which causes dislocation however with the capsule being torn there was extra space anteriorly which allowed for the hip to sublux anteriorly.Doi: (b)(6) 2012 - dor: (b)(6) 2012 (left hip, second revision) please see (b)(6), for the first revision.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #: (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASPHERE M SPEC 12/14 40 +1.5
Common Device Name
ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9207717
Report Number1818910-2019-110650
Device Sequence Number19591043
Product Code JDI
UDI-Device Identifier10603295032977
UDI-Public10603295032977
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K082585
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2012
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/30/2016
Device Catalogue Number136505100
Device Lot Number3345786
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/27/2019
Supplement Date Received by Manufacturer11/18/2019
Initial Report FDA Received Date10/18/2019
Supplement Report FDA Received Date11/18/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age76 YR
-
-