| Model Number |
7143 |
| Medical Device Problem Code |
Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/30/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that catheter entrapment occurred.Two 2.50mm x 15mm emerge balloon catheters were used together with 2 different bmw wires consecutively.However, the both devices got stuck with the wires and would not move.The procedure was completed with non-bsc devices.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of an emerge balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There was blood in the guidewire lumen.The balloon was tightly folded.The device was functionally tested with a.014" guidewire.The guidewire was unable to advance due to the inner shaft being buckled for 14mm in the strain relief.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
It was reported that catheter entrapment occurred.Two 2.50mm x 15mm emerge balloon catheters were used together with 2 different bmw wires consecutively.However, the both devices got stuck with the wires and would not move.The procedure was completed with non-bsc devices.No patient complications were reported.
|
| |
|