• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC RADIAL ARTERY PRESSURE MONITORING SET; DQO CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC RADIAL ARTERY PRESSURE MONITORING SET; DQO CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Model Number N/A
Medical Device Problem Code Delivered as Unsterile Product (1421)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 09/26/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).Occupation: unknown.Pma/510(k) #: not exempt, pre-amendment.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported that a hair was found inside of the primary packaging of a radial artery pressure monitoring set during an inspection.No patient contact was made.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Investigation - evaluation: a review of the documentation including the complaint history, device history record (dhr) and quality control, as well as a visual inspection of photos of the device was conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, photos of the device were provided from the complainant.The complaint was able to be confirmed based on customer testimony and the provided photos, which revealed a small, unidentified particle inside of the sealed packaging.Consequently, there is evidence that this device was manufactured out of specification.Additionally, a document-based investigation evaluation was performed.It was concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the device history record (dhr) for the complaint lot (9585859) revealed no related non-conformances.A database search found this to be the only complaint associated with the complaint device lot.Therefore, there is no evidence of additional non-conforming material or products from this lot in house or in the field.Based on the information provided, no product returned, and the results of our investigation, a definitive root cause can be traced back to a quality control deficiency during manufacturing.The appropriate personnel have been notified.Per the quality engineering risk assessment no further action is required.Cook will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RADIAL ARTERY PRESSURE MONITORING SET
Common Device Name
DQO CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key9208776
Report Number1820334-2019-02645
Device Sequence Number19619612
Product Code DQO
UDI-Device Identifier00827002019089
UDI-Public(01)00827002019089(17)220312(10)9585859
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date03/12/2022
Device Model NumberN/A
Device Catalogue NumberC-PMS-300-RA
Device Lot Number9585859
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/04/2019
Supplement Date Received by Manufacturer11/20/2019
Initial Report FDA Received Date10/18/2019
Supplement Report FDA Received Date11/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-