| Model Number |
10601 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/30/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device is a combination product.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.The tightly stenosed target lesion with a 90-degree bend was located in the circumflex artery.A 2.50 x 12mm synergy drug-eluting stent was advanced multiple times through a very calcified left main coronary artery.However, during the third or fourth time, the shaft snapped.A trapping device was used to trap and pull out the entire system.The procedure was completed with a different device.No patient complications were reported and the patient was fine.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.The tightly stenosed target lesion with a 90-degree bend was located in the circumflex artery.A 2.50 x 12mm synergy drug-eluting stent was advanced multiple times through a very calcified left main coronary artery.However, during the third or fourth time, the shaft snapped.A trapping device was used to trap and pull out the entire system.The procedure was completed with a different device.No patient complications were reported and the patient was fine.
|
| |
|
Additional Manufacturer Narrative
|
|
Device is a combination product.Device evaluated by mfr.: synergy ii us mr 2.50 x 12 mm stent delivery system was returned for analysis broken in 2 pieces.A visual examination of the stent found no issues.There was no sign of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set between the proximal and distal markerbands.The crimped stent outer diameter was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed signs of damage on the distal end of the tip.A visual and tactile examination of the hypotube shaft found a break situated at 76.9 cm distal to the distal end of the strain relief as well as multiple hypotube kinks.A fracture in the laser cut region with the corewire exposed was also noted and measured at 38.6cm proximal to the distal tip.A visual and tactile examination of the outer and mid-shaft section and a visual examination of the inner lumen found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.
|
| |
|
Search Alerts/Recalls
|