| Catalog Number |
465352CB2 |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/18/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The suspect device has been returned for evaluation.A follow up will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The account alleges that during a peripheral vascular interventional procedure, the tip of the catheter detached within the patient.The physician acquired left retrograde femoral arterial access and had negotiated the patient's common iliac bifurcation with a guidewire and catheter.During catheter manipulations within the contralateral external iliac artery post-stent placement, the catheter became stuck within the patient's artery.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device has been returned for evaluation.The product was examined visually.The complaint is confirmed.The root cause was attributed to excessive force being applied to the device during use.The device history record was reviewed and no exception documents were found.A search of the complaint database was performed and no similar complaints for this lot number were found.
|
| |
|