• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MASIMO - 52 DISCOVERY RD SET NEOPT-500; OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 52 DISCOVERY RD SET NEOPT-500; OXIMETER Back to Search Results
Model Number 4005
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/23/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional manufacturing narrative: attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.(b)(6).
 
Event or Problem Description
The customer reported the device provided low spo2 readings in the nicu.No patient impact or consequences were reported.
 
Event or Problem Description
The customer reported the device provided low spo2 readings in the nicu.No patient impact or consequences were reported.
 
Additional Manufacturer Narrative
Additional manufacturing narrative: the returned sensor was evaluated.During evaluation the sensor passed all visual and functional testing.  during simulation testing, the sensor passed manual and preset conditions and provided accurate measurements.The sensor was determined to be functioning as designed.   initial reporter zip code is entered as "00000" to meet the maximum allowable characters.Initial reporter zip code is (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RD SET NEOPT-500
Common Device Name
OXIMETER
Manufacturer (Section D)
MASIMO - 52 DISCOVERY
52 discovery
irvine CA 92618
MDR Report Key9209729
Report Number3011353843-2019-00093
Device Sequence Number3146639
Product Code DQA
UDI-Device Identifier10843997009895
UDI-Public10843997009895
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K042536
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 09/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/01/2022
Device Model Number4005
Device Catalogue Number4005
Device Lot NumberE19BZP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/22/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 09/24/2019
Supplement Date Received by Manufacturer11/08/2019
Initial Report FDA Received Date10/18/2019
Supplement Report FDA Received Date12/02/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Weight1
-
-