| Model Number |
4005 |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/23/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional manufacturing narrative: attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.(b)(6).
|
| |
|
Event or Problem Description
|
|
The customer reported the device provided low spo2 readings in the nicu.No patient impact or consequences were reported.
|
| |
|
Event or Problem Description
|
|
The customer reported the device provided low spo2 readings in the nicu.No patient impact or consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacturing narrative: the returned sensor was evaluated.During evaluation the sensor passed all visual and functional testing. during simulation testing, the sensor passed manual and preset conditions and provided accurate measurements.The sensor was determined to be functioning as designed. initial reporter zip code is entered as "00000" to meet the maximum allowable characters.Initial reporter zip code is (b)(6).
|
| |
|