• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC DEPTH GAUGE FOR 2.7MM & SMALL SCREWS; GAUGE, DEPTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC DEPTH GAUGE FOR 2.7MM & SMALL SCREWS; GAUGE, DEPTH Back to Search Results
Model Number 319.04
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Without a lot number the device history records review could not be completed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2019, the slender metal extension of a depth gauge that goes down into the drill hole is breaking off the handle.It was discovered during cleanup.There was no patient involvement.Concomitant medical product: unknown handle (part # unknown, lot # unknown, quantity 1).This report is for 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: a1, a2, a3, a4: device used in a veterinary case - no patient information will be reported.H3, h6: visual inspection: the depth gauge for 2.7 mm and small screws was returned and received at us customer quality (cq).Upon visual inspection, it was observed that the needle component was broken from slider and returned to cq.No other issues were identified with the returned components of the device.Service & repair evaluation: the customer reported the device was broken.The repair technician reported the device tip was broken.Tip broken is the reason for repair.The item is not repairable per the inspection sheet.The cause of the issue is unknown.The item will be forwarded to customer quality.The evaluation was confirmed.Dimensional inspection: no dimensional inspection can be performed due to post-manufacturing damage.Furthermore, the complaint relevant dimensions cannot be checked for dimensional accuracy, because of the design of the device.Document/specification review based on the date of manufacture, the following drawings, reflecting the current and manufactured revision were reviewed.Slider assembly; depth gauge for 2.7 mm to 4.0 mm screws.Investigation conclusion: the complaint condition is confirmed for the depth gauge for 2.7 mm and small screws.There is no indication that a design or manufacturing issue has caused the issue and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEPTH GAUGE FOR 2.7MM & SMALL SCREWS
Common Device Name
GAUGE, DEPTH
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key9209904
Report Number2939274-2019-61661
Device Sequence Number18944396
Product Code HTJ
UDI-Device Identifier10886982189974
UDI-Public(01)10886982189974
Combination Product (Y/N)N
Initial Reporter CountryWA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number319.04
Device Catalogue Number319.04
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/14/2019
Initial Date Received by Manufacturer 10/02/2019
Supplement Date Received by Manufacturer11/22/2019
Initial Report FDA Received Date10/18/2019
Supplement Report FDA Received Date11/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK - HANDLES: TRAUMA.
-
-