DEPUY IRELAND 9616671 ACTIS BROACH SZ 4; BROACHES Back to Search Results
Catalog Number 201001040
Device Problems Difficult to Remove (1528); Dull, Blunt (2407); Connection Problem (2900); Scratched Material (3020)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/21/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: examination of the returned instrument found the post to be scratched and the cutting teeth to be dull.The root cause is attributed to misuse and wear.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Event Description
After investigation the instrument was determined to be nicked.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #
=
> (b)(4).Investigation summary
=
> examination of the returned instrument found the post to be scratched and the cutting teeth to be dull.The root cause is attributed to misuse and wear.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot
=
> null.Device history batch
=
> null.Device history review
=
> null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIS BROACH SZ 4
Type of Device
BROACHES
Manufacturer (Section D)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI 
MDR Report Key9210930
MDR Text Key184989212
Report Number1818910-2019-110802
Device Sequence Number1
Product Code KWL
UDI-Device Identifier10603295393665
UDI-Public10603295393665
Combination Product (y/n)N
PMA/PMN Number
K150862
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 10/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/18/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number201001040
Device Lot NumberPG274007
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2019
Date Manufacturer Received10/17/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-