• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DRÄGERWERK AG & CO. KGAA PRIMUS; ANESTHESIA UNITS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DRÄGERWERK AG & CO. KGAA PRIMUS; ANESTHESIA UNITS Back to Search Results
Catalog Number 8603800
Device Problems Failure to Deliver (2338); No Pressure (2994)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/23/2019
Event Type  malfunction  
Manufacturer Narrative
The investigation has just started; results will be provided in a follow up-report.
 
Event Description
It was reported that there was a ventilator failure during use.Manual ventilation remained possible.There was no patient injury reported.
 
Event Description
Please refer to the initial-report.
 
Manufacturer Narrative
Based on the available log file, the reported case could be reconstructed.It was found that during the case in question, a failure of the inspiratory pressure sensor of the vgc (ventilation and gas controller) was detected by the device.In consequence, the ventilator initiated an autonomous shutdown while changing mode to man/spont accompanied by the corresponding "ventilator fail" alarm as specified.In this case, manual ventilation as well as monitoring function remains unaffected.On-site, in follow-up of the event, the dispatched fse replaced the affected pressure sensor as a precautionary measure.The device was successfully tested afterwards and was returned to use without further problems reported.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIMUS
Type of Device
ANESTHESIA UNITS
Manufacturer (Section D)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck
GM 
MDR Report Key9212702
MDR Text Key186751430
Report Number9611500-2019-00341
Device Sequence Number1
Product Code BSZ
UDI-Device Identifier04048675041436
UDI-Public(01)04048675041436(11)190311(17)190808(93)8603800-94
Combination Product (y/n)N
PMA/PMN Number
K042607
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 11/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/21/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8603800
Was Device Available for Evaluation? Yes
Date Manufacturer Received11/05/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-