| Model Number |
A0820-07 |
| Medical Device Problem Code |
Compatibility Problem (2960)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The manufacturer's investigation is on-going and further information will be provided once the investigation has completed.
|
| |
|
Event or Problem Description
|
|
Mr fiducial box use error.
|
| |
|
Additional Manufacturer Narrative
|
|
H6 updated.H10 updated: following start up at a site elekta were informed about a user error relating to the mr fiducial box.Oil had been used instead of cuso4 as stipulated by the instruction for use.Use of oil is known from literature to cause shifts in the stereotactic reference.The issue has been solved at site with delivery of a new mr indicator box.Elekta has contacted the site to offer help in investigating the consequences of the use of oil.The site has not communicated to elekta that they would accept our help and has not informed elekta that they have reported any patient mistreatment as a consequence of the use of oil.
|
| |
|
Event or Problem Description
|
|
Mr fiducial box use error.
|
| |
|