It was reported an unknown patient required a quick-core coaxial biopsy needle set for an unknown procedure.Prior to patient contact, the operator noticed the coaxial needle assembly was "difficult to unscrew" and the "set could not be separated." as reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
|
|
Concomitant medical product received on: 28oct2019.Investigation ¿ evaluation.It was reported that the coaxial needle assembly within a quick-core coaxial biopsy needle set was difficult to unscrew.Further communication with the user facility clarified that the device did not make patient contact.The patient did not experience any adverse effects.A review of the complaint history, device history record, drawing, instructions for use (ifu), quality control, and specifications of the device, as well as a visual inspection, functional test, and dimensional verification, were conducted during the investigation.One opened coaxial needle subassembly (dtn) and one unopened biopsy needle (qc) were returned.No surface damage was noted on the dtn, but there was some slight biological matter.The dtn was difficult to unscrew at first, but was eventually able to be unscrewed by hand.Retightening the device replicates the failure partially; the device is again difficult to unscrew at first, but was able to be unscrewed by hand.Dimensions such as stylet outer diameter and cannula inner diameter were measured within specification.Additionally, a document based investigation evaluation was performed.The risks associated with these devices are acceptable when weighed against the benefits.Although inspection steps are in place related to this failure mode, a project was opened to further investigate/address this issue.Product labeling was also reviewed.Instructions for use are packaged with this device.Relevant sections include: intended use: quick-core biopsy needles are intended for soft tissue biopsy.The product is intended for use by physicians trained and experienced in diagnostic and interventional techniques.Standard techniques for soft tissue biopsy should be employed.How supplied: upon removal from package, inspect the product to ensure no damage has occurred.The device history record (dhr) for the complaint lot showed no nonconformances.Relevant coaxial needle subassembly lot showed no related nonconformances.A database search revealed one other complaint from the lot at the time of investigation.This complaint was reported from the same distributor and concerns the same failure mode.There is no evidence to suggest nonconforming product exists in house or in the field.It is possible that the device was screwed too tightly during quality control, but this cannot be definitively confirmed.Another unknown factor may be contributing to the dtn becoming unable to be unscrewed.Based on the information provided, examination of the returned product, and the results of the investigation, a definitive cause could not be established.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the quality engineering risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|