• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH MODULAR DUAL MOBILITY INSERT; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH MODULAR DUAL MOBILITY INSERT; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS Back to Search Results
Catalog Number 626-00-46F
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Injury (2348); Joint Dislocation (2374)
Date of Event 09/23/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
Hip revision due to disassociated mdm inner bearing from outer bearing.
 
Additional Manufacturer Narrative
An event regarding dislocation involving an mdm insert was reported.The event was not confirmed.Method & results: device evaluation and results: not performed as product was not returned.Clinician review: no medical records were received for review with a clinical consultant.Device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies.Complaint history review: there have been no other similar events for the reported lot.Conclusion: the exact cause of the event could not be determined because insufficient information was provided.Additional information including operative reports, progress notes, x-rays and return of the device are needed to fully investigate the event.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Hip revision due to disassociated mdm inner bearing from outer bearing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MODULAR DUAL MOBILITY INSERT
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key9213557
Report Number0002249697-2019-03562
Device Sequence Number10899277
Product Code LZO
UDI-Device Identifier04546540641243
UDI-Public04546540641243
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K103233
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2023
Device Catalogue Number626-00-46F
Device Lot Number63804801
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/23/2019
Supplement Date Received by Manufacturer11/28/2019
Initial Report FDA Received Date10/21/2019
Supplement Report FDA Received Date12/26/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age55 YR
Patient Weight102
-
-