STRYKER ORTHOPAEDICS-MAHWAH UNIVERSAL IMPACTOR/POSTIONER; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 2101-0200
Device Problems Loss of or Failure to Bond (1068); Mechanical Problem (1384)
Patient Problems No Consequences Or Impact To Patient (2199); No Patient Involvement (2645)
Event Date 09/23/2019
Event Type  malfunction  
Manufacturer Narrative
If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
 
Event Description
It was reported that upon routine inspection of hospital inventory, the junction of the metal rod and the handle of the cup inserter was loose.No associated procedure.
 
Manufacturer Narrative
An event regarding damage involving cutting edge impactor was reported.The reported event was confirmed.Conclusion: the device was discovered during inspection.The reported device was returned for evaluation.A material analysis has been performed.The report concluded: examination of the returned device with engineer indicated that the device is severely worn due to in service use.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.There was no surgical procedure associated with the reported event.This event meets the definition of preventive maintenance.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event Description
It was reported that upon routine inspection of hospital inventory, the junction of the metal rod and the handle of the cup inserter was loose.No associated procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL IMPACTOR/POSTIONER
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key9214835
MDR Text Key187599683
Report Number0002249697-2019-03567
Device Sequence Number1
Product Code LPH
UDI-Device Identifier07613327211511
UDI-Public07613327211511
Combination Product (y/n)N
PMA/PMN Number
K161569
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 01/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/21/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2101-0200
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2019
Date Manufacturer Received12/12/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-