PENUMBRA, INC. PENUMBRA COIL 400; HCG, KRD Back to Search Results
Catalog Number 4006C0306
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/04/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report numbers: 3005168196-2019-01961; 3005168196-2019-01963.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the inferior mesenteric artery (ima) using penumbra coil 400s (pc400s), a pod4 and a penumbra coil 400 detachment handle (handle).During the procedure, while attempting to advance a pc400 into the target vessel using a lantern delivery microcatheter (lantern), the microcatheter kicked out of place.Therefore, the physician removed the pc400 and repositioned the lantern.While attempting to advance the same pc400 through the lantern, the physician experienced strong resistance and therefore, the pc400 was removed.It was reported that flushing the microcatheter did not resolve the resistance issue.Next, the physician advanced a pod4 and attempted to detach the coil using a handle; however, the pod4 failed to detach and, therefore, it was removed.It was also reported that the pod4 alignment zone was not divided in half.The procedure was completed using additional pc400s with the same handle and lantern.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA COIL 400
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key9214988
Report Number3005168196-2019-01962
Device Sequence Number2524336
Product Code HCG
UDI-Device Identifier00814548011837
UDI-Public00814548011837
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/23/2019
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/27/2026
Device Catalogue Number4006C0306
Device Lot NumberF83683
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/23/2019
Initial Report FDA Received Date10/21/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age81 YR
Date Report Sent to FDA01/01/2005
-
-