| Catalog Number |
03L79-31 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/26/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
All available patient information was included.Additional patient details are not available.No customer returns were available for evaluation.A review of ticket trending for architect calcium reagent product lot number (61157un18) did identify one additional complaint that was similar to this issue.In that case, no deficiency was identified.The trend review by the product list number found no trends related to this issue.Worldwide field data was reviewed to evaluate the historical performance of lot 61157un18.The patient median result for the lot was comparable with all other lots in the field and within established limits, confirming no systemic issue.Labeling was reviewed and found to adequately address the issue under review.No product deficiency was identified.This issue was previously reported under mdr number 1415939-2019-00170 under the same suspect medical device but an incorrect manufacturer name, city and state.
|
| |
|
Event or Problem Description
|
|
The customer reported a falsely decreased calcium result generated on the architect c8000 analyzer on one patient.Results provided: sid (b)(6) on (b)(6) 2019 = 2.3 mg/dl / repeated same sample: sid (b)(6) on (b)(6) 2019 = 9.6 mg/dl.(normal range: 8.5-10.5 mg/dl) there was no impact to patient management reported.
|
| |
|
Search Alerts/Recalls
|