| Catalog Number |
10636672 |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/23/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer stated that repeat testing was performed to confirm correct results and a corrected report was issued.The customer declined to send any materials for further investigation.Proper technique and maintenance were reviewed with the customer as well as possible causes of a falsely depressed leukocyte result.The customer had not lubricated the push bar as often as recommended and are operating at a slightly less than recommended temperature.Since they installed the new fixed platform and lubricated the push-bar rail, they have not had any further discrepancies.The instrument is operational.
|
| |
|
Event or Problem Description
|
|
The customer reported discrepant leukocyte results on the clinitek advantus when compared to the microscopic examination of the urine sediment.There is no report of injury due to this event.
|
| |
|