MAUDE Adverse Event Report: COOK ENDOSCOPY QUANTUM TTC BILIARY BALLOON DILATOR; FGE, BILIARY CATHETER AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK ENDOSCOPY QUANTUM TTC BILIARY BALLOON DILATOR; FGE, BILIARY CATHETER AND ACCESSORIES Back to Search Results
Model Number G22655
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/18/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Non-healthcare professional.Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.It is unknown if negative pressure and lubrication were applied to the balloon prior to advancement.The instructions for use direct the user: "prior to insertion of balloon dilator, negative pressure is mandatory to maintain balloon deflation." the application of negative pressure will aspirate all residual air from the balloon and ease endoscopic advancement.Negative pressure will also aid in balloon preservation and optimize balloon performance.The instructions for use advise the user: "apply a water soluble lubricant to balloon to allow easier passage through accessory channel." this will aid in balloon advancement and preservation.Another possible contributing factor is if the device comes in contact with a sharp object or burr in the endoscope accessory channel.Another possible contributing factor is using a compromised inflation device to inflate the balloon.If the pressure reading of the inflation device is inaccurate, this could contribute to over inflation, possibly resulting in damage to the balloon material.Prior to distribution, all quantum ttc biliary balloon dilators are subjected to a leak test to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event or Problem Description
During an endoscopic retrograde cholangiopancreatography (ercp), the physician used a cook quantum ttc biliary balloon dilator.The user advanced the device through wire guide to duodenal papilla and inflated the balloon.The user found a leaking issue.The user retracted the device and changed another same device to complete the procedure.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Continued from section e3: non-healthcare professional.Investigation evaluation: our laboratory evaluation of the product said to be involved confirmed the report.A functional test was performed on the returned device; a cook quantum biliary inflation device (qbid-1) was filled with water and attached to the balloon inflation port, and negative pressure was applied to the balloon.After applying negative pressure, the balloon was inflated.The balloon would not hold pressure, and water was leaking from a pinhole on the proximal side of the balloon.An examination of the gold bands near the distal and proximal ends of the balloon showed no roughness.A discrepancy or anomaly that could have contributed to the reported event was not observed during our laboratory analysis of the returned product.The device history record for all possible lot numbers said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.Another possible contributing factor is if the device comes in contact with a sharp object or burr in the endoscope accessory channel.Another possible contributing factor is using a compromised inflation device to inflate the balloon.If the pressure reading of the inflation device is inaccurate, this could contribute to over inflation, possibly resulting in damage to the balloon material.Prior to distribution, all quantum ttc biliary balloon dilators are subjected to a leak test to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event or Problem Description
During an endoscopic retrograde cholangiopancreatography (ercp), the physician used a cook quantum ttc biliary balloon dilator.The user advanced the device through wire guide to duodenal papilla and inflated the balloon.The user found a leaking issue.The user retracted the device and changed another same device to complete the procedure.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM TTC BILIARY BALLOON DILATOR
Common Device Name
FGE, BILIARY CATHETER AND ACCESSORIES
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
MDR Report Key9216034
Report Number1037905-2019-00622
Device Sequence Number2181319
Product Code FGE
UDI-Device Identifier00827002226555
UDI-Public(01)00827002226555(17)211017(10)W4133496
Combination Product (Y/N)N
Initial Reporter CountryCN
PMA/510(K) Number
K171223
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/17/2021
Device Model NumberG22655
Device Catalogue NumberQBD-6X3
Device Lot NumberW4133496
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/28/2019
Initial Date Received by Manufacturer 09/26/2019
Supplement Date Received by Manufacturer10/28/2019
Initial Report FDA Received Date10/21/2019
Supplement Report FDA Received Date11/20/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ENDOSCOPE, UNKNOWN MAKE OR MODEL; OLYMPUS ENDOSCOPE, UNKNOWN MODEL
Patient Age62 YR
-
-