| Model Number |
375-544-000 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/24/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device was not returned to stryker sustainability solutions for evaluation.The device is being held by the facility.As the device was not returned for evaluation, inspection was unable to be performed.A review of the dhr for the reported lot number supports that the device met all inspection and test criteria prior to release from stryker.The reported event could be attributed to user applied too much force to the device or the arthroscopic shaver may have come into contact with staples, clips or another metal object, resulting in damage to the blade.The instructions for use (ifu) state: before beginning the procedure, verify compatibility of all instruments and accessories.Plug in and set up the generator according to the instructions in the manufacturers¿ manual.Select an arthroscopic shaver with size, blade, and function most appropriate for the procedure.Inspect the instrument for overall condition and physical integrity.Do not use the instrument if any damage is noted.Returnthe instrument and packaging to stryker sustainability solutions if it is not in acceptable condition for surgery.Careful handling of the instrument is necessary to avoid damage or breakage as a result of excessive force.Do not apply excessive pressure or ¿side-load¿ the blade during use.Side-loading does not improve the performance of the instrument, can dull the blade, and/or produce metal particulates.Do not allow the arthroscopic shaver to come into contact with staples, clips or any metal object to avoid damage to the blade and possible patient injury.The tip of the bur or cutter must be irrigated periodically (general recommendation: once a minute) to cool the blade and prevent excised tissues from accumulating.Do not run the instrument without appropriate suction for the duration of the process.The reported event will continue to be monitored through post-market surveillance.Should the device become available for return, the investigation will be reopened.
|
| |
|
Event or Problem Description
|
|
It was reported the tip of the shaver broke off in the patient during the procedure.There was no patient injury and extended procedure time reported was 20-30 minutes to remove.Some additional anesthesia was given to keep the patient sedated.The c arm was brought in to locate the piece.The piece was removed with a grasper.These are commonly used devices that are readily available.
|
| |
|
Additional Manufacturer Narrative
|
|
On 11/18/2020, stryker sustainability solutions became aware that mdr section b2.(outcomes attributed to ae) was not completed for this mdr.This supplemental mdr serves to provide this information.The reported event will continue to be monitored through post market surveillance.
|
| |
|
Event or Problem Description
|
|
It was reported the tip of the shaver broke off in the patient during the procedure.There was no patient injury and extended procedure time reported was 20-30 minutes to remove.Some additional anesthesia was given to keep the patient sedated.The c arm was brought in to locate the piece.The piece was removed with a grasper.These are commonly used devices that are readily available.
|
| |
|