DAVIS & GECK CARIBE LTD BIOSYN; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID Back to Search Results
Model Number UM-204
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/03/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter during a veterinary procedure, the needle and thread of the device broke right after they attempt to use the device.Another device was used to resolve the issue.There was no patient harm.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of two devices.The visual inspection of the samples noted two sutures and two needles.The needles were disengaged from the sutures, and no other breaks were noted in the suture or needles.Upon microscopic inspection of the needles, normal crimp and swage marks were noted.As no sealed samples were returned, a needle attachment test could not be performed.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOSYN
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo,ct 0101
DO  0101
MDR Report Key9216721
Report Number9612501-2019-02043
Device Sequence Number19372441
Product Code GAM
UDI-Device Identifier10884521033368
UDI-Public10884521033368
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K000037
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,user faci
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2023
Device Model NumberUM-204
Device Catalogue NumberUM-204
Device Lot NumberD9A2157Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/01/2019
Initial Date Received by Manufacturer 10/03/2019
Supplement Date Received by Manufacturer12/23/2019
Initial Report FDA Received Date10/21/2019
Supplement Report FDA Received Date12/27/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-