Model Number BEQ-HLS 7050 USA#HLS SET ADVANCED 7.0 |
Device Problem
Pressure Problem (3012)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 09/28/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
A follow up medwatch will be submitted when additional information becomes available.
|
|
Event Description
|
I received a call from (b)(6), ccp who reported that they had a patient on an hls set and the p art went from around 110-120 mmhg down to 20 mmhg for no known reason.The p interior remained the same as well as the flow, so she believed it was a faulty pressure.She decided to change the circuit to a new hls set and the pressures were normal with the new disposable.She agreed to keep the disposable to send in for inspection.Complaint id: (b)(4).
|
|
Event Description
|
Complaint id: (b)(4).
|
|
Manufacturer Narrative
|
The suspected hls module was investigated by the manufacturer.The visible inspection did not show any defects on the internal sensors or the oxygenator.A tightness test according to lv 201 and a pressure test with a connected cardiohelp was performed.No abnormalities were discovered.No root cause could be determined.Further a device history review was performed and there were no references found, which are indicating a nonconformance of the product in question.Thus the failure could not be confirmed.The occurrence rate regarding the above complaint is below the acceptance rate.Thus, no remedial action required.
|
|