• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH HLM TUBING SET W/BIOLINE COATING; TUBING, PUMP, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH HLM TUBING SET W/BIOLINE COATING; TUBING, PUMP, CARDIOPULMONARY BYPASS Back to Search Results
Model Number BEQ-HLS 7050 USA#HLS SET ADVANCED 7.0
Device Problem Pressure Problem (3012)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/28/2019
Event Type  malfunction  
Manufacturer Narrative
A follow up medwatch will be submitted when additional information becomes available.
 
Event Description
I received a call from (b)(6), ccp who reported that they had a patient on an hls set and the p art went from around 110-120 mmhg down to 20 mmhg for no known reason.The p interior remained the same as well as the flow, so she believed it was a faulty pressure.She decided to change the circuit to a new hls set and the pressures were normal with the new disposable.She agreed to keep the disposable to send in for inspection.Complaint id: (b)(4).
 
Event Description
Complaint id: (b)(4).
 
Manufacturer Narrative
The suspected hls module was investigated by the manufacturer.The visible inspection did not show any defects on the internal sensors or the oxygenator.A tightness test according to lv 201 and a pressure test with a connected cardiohelp was performed.No abnormalities were discovered.No root cause could be determined.Further a device history review was performed and there were no references found, which are indicating a nonconformance of the product in question.Thus the failure could not be confirmed.The occurrence rate regarding the above complaint is below the acceptance rate.Thus, no remedial action required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HLM TUBING SET W/BIOLINE COATING
Type of Device
TUBING, PUMP, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
MDR Report Key9217477
MDR Text Key200640827
Report Number8010762-2019-00335
Device Sequence Number1
Product Code DWE
Combination Product (y/n)N
PMA/PMN Number
K080592
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign,health profe
Type of Report Initial,Followup
Report Date 11/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/19/2021
Device Model NumberBEQ-HLS 7050 USA#HLS SET ADVANCED 7.0
Device Catalogue Number70105.2794
Device Lot Number70129850
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2019
Initial Date Manufacturer Received 09/28/2019
Initial Date FDA Received10/22/2019
Supplement Dates Manufacturer Received09/28/2019
Supplement Dates FDA Received11/18/2019
Patient Sequence Number1
-
-