• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W.L. GORE & ASSOCIATES GORE MYCROMESH® PLUS BIOMATERIAL; MESH, SURGICAL, POLYMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W.L. GORE & ASSOCIATES GORE MYCROMESH® PLUS BIOMATERIAL; MESH, SURGICAL, POLYMERIC Back to Search Results
Model Number 1MYMP01
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Obstruction/Occlusion (2422)
Date of Event 09/14/2006
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(6).It should be noted that the instructions for gore® mycromesh® plus biomaterial use includes warnings and addresses the following adverse reactions among others: ¿possible adverse reactions with the use of any tissue deficiency prosthesis may include, but are not limited to, contamination, infection, inflammation, adhesion, fistula formation, seroma formation, hematoma, and recurrence.¿.
 
Event or Problem Description
It was reported to gore that the patient underwent open umbilical hernia repair on (b)(6) 2004 and (b)(6) 2007 whereby a gore® mycromesh® plus biomaterial was implanted.The complaint alleges that on (b)(6) 2006, an additional procedure occurred whereby the gore device was explanted.The complaint alleges that on (b)(6) 2017, an additional procedure occurred whereby the gore device was explanted.It was reported the patient alleges the following injuries: removal of mesh and small bowel obstruction ((b)(6) 2006); removal of mesh and infection ((b)(6) 2017).Additional event specific information was not provided.
 
Additional Manufacturer Narrative
H6: updated results code.Conclusion code remains unchanged.
 
Additional Manufacturer Narrative
A4: added patient weight.B7: added medical history.H6: updated results code 2 for sterilization evaluation.Conclusion code remains unchanged.H10/11: added medical record information.Additional details regarding the patient's clinical course were ascertained from a review of medical records and are as follows: please see the attached file for information regarding the clinical course obtained from the medical records submitted on 10/25/19.A potential relationship, if any, between the alleged injuries or complications and the gore device has not been established at this time based on available information.It should be noted that the instructions for gore® mycromesh® plus biomaterial use includes warnings and addresses the following adverse reactions among others: ¿possible adverse reactions with the use of any tissue deficiency prosthesis may include, but are not limited to, contamination, infection, inflammation, adhesion, fistula formation, seroma formation, hematoma, and recurrence.¿ w.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Additional Manufacturer Narrative
H6: health effect impact code: f1903: device explantation.H6: medical device component: g04088: membrane.The investigation has been completed.Based upon the totality of the information received over the course of the investigation and reported by gore in the previously submitted medical record narratives the following conclusions have been reached.As previously reported, all pertinent medical records requested may not have been received.In the absence of additional information or medical records from the complainant, this event file will be closed with the information provided.It should be noted that the gore® mycromesh® plus biomaterial instructions for use include warnings and addresses the following adverse reactions among others: ¿possible adverse reactions with the use of any tissue deficiency prosthesis may include, but are not limited to, contamination, infection, inflammation, adhesion, fistula formation, seroma formation, hematoma, and recurrence.¿ as with any surgical procedure, there are always risks of complications for surgical repair of hernias and soft tissue deficiencies, with or without mesh.These may include but are not limited to, adhesions and related harms, bleeding, bowel obstruction, compromised device biocompatibility, contamination which may lead to patient harms, device damage, dysphagia, erosion or extrusion and related harms, exposure or protrusion and related harms, fever, fistula, gerd recurrence, defect recurrence and related harms, ileus, increased procedure time and related harms, irritation or inflammation, infection, mesh migration, mesh contraction, pain, paresthesia, perforation, revision/re-intervention, seroma or hematoma and related harms, tissue ischemia, wound complications and wound dehiscence and additional intervention including surgery.Many of the potential complications are associated with the patient¿s underlying disease progression, co-morbidities, additional medical history and/or other surgical procedures.The above inherent risks are typically detailed in standard informed consent documents.The device was not able to be returned to gore for evaluation; therefore, a direct product analysis could not be conducted.Review of the manufacturing and sterilization records verified that the lot met all pre-release specifications.Section c1: name: plus antimicrobial product coating; manufacturer/compounder: w.L.Gore & associates, inc.; lot number: 02076073.Additional manufacturer narrative: the plus antimicrobial product coating contains silver carbonate [approximately 800 micrograms per cubic centimeter of product (g/cm3)], and chlorhexidine diacetate [approximately 1600 micrograms per cubic centimeter of product (g/cm3)].W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE MYCROMESH® PLUS BIOMATERIAL
Common Device Name
MESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
W.L. GORE & ASSOCIATES
flagstaff AZ
Manufacturer (Section G)
MEDICAL WOODY MOUNTAIN B/P
3750 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
rena gros
1500 n. 4th street
9285263030
MDR Report Key9217568
Report Number2017233-2019-01063
Device Sequence Number2166615
Product Code FTL
UDI-Device Identifier00733132601615
UDI-Public00733132601615
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K063435
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2004
Device Explanted Year2006
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 06/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Expiration Date03/15/2005
Device Model Number1MYMP01
Device Catalogue Number1MYMP01
Device Lot Number02076073
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Not provided
Not provided
Initial Report FDA Received Date10/22/2019
Supplement Report FDA Received Date11/11/2019
12/13/2019
06/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/16/2003
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age70 YR
Patient SexMale
Patient Weight93 KG
-
-