| Model Number |
956-623 |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Fail-Safe Did Not Operate (4046)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); Needle Stick/Puncture (2462)
|
| Date of Event |
10/07/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The affected device is not available and has not been returned to radiometer.The date of event stated is best approximation.Importers report number: (b)(4).
|
| |
|
Event or Problem Description
|
|
A respiratory therapist had a needle stick from safepico blood sampler.After normal blood sampling procedure the therapist could not activate the safety shield of the sampler, which ultimately led to a finger stick.The therapist stated that she has experienced several blood samplers where this safety mechanisms fails to work.The therapist was following tested in the emergency department after the needle stick.
|
| |
|
Additional Manufacturer Narrative
|
|
In this follow-up report #1 the patient code(s) and device code(s) has been updated in field h6 to reflect the codes used in the associated initial importers report.
|
| |
|
Additional Manufacturer Narrative
|
|
The defective sampler was not returned to radiometer and could therefore not be examined.Reference samples were checked under visual inspection.The checked samples were all in specification with no issues observed.Additionally mechanical test were performed.There were no problems with safeguard opening observed.All the tested samples were with proper functionality.Since the defective sampler was not returned and no issues in reference samples were observed, the root cause could not be identified.
|
| |
|