• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MFG DC BOULDER COOL-TIP; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG DC BOULDER COOL-TIP; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number ACT2030
Medical Device Problem Code No Flow (2991)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/04/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a procedure, the water pipe of system pump would not start up to capture water to device.Another device was used to complete the procedure.There was no patient injury.
 
Additional Manufacturer Narrative
Additional info: a5b (asian).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Evaluation summary: one device was received for evaluation.The returned product did not meet specification as received.Visual inspection found no defects.The reported condition was confirmed.The investigation found the sample did not allow water to circulate through.The tubing was not blocked.The needle was removed and fiber was found at the tip of the hypotube/thermocouple assembly.Engineering and the supplier have examined all the manufacturing processes for the needles.Nothing in the supplier or manufacturing process would contribute to the fibers found.The investigation identified the root cause of the reported event to be user error.The customer is advised to use sterile saline for cooling.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOL-TIP
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
MDR Report Key9217830
Report Number1717344-2019-01380
Device Sequence Number2183420
Product Code GEI
UDI-Device Identifier10884524001227
UDI-Public10884524001227
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K973297
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/29/2022
Device Model NumberACT2030
Device Catalogue NumberACT2030
Device Lot Number90730053X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2020
Initial Date Received by Manufacturer 10/04/2019
Supplement Date Received by Manufacturer10/29/2019
01/22/2020
Initial Report FDA Received Date10/22/2019
Supplement Report FDA Received Date11/13/2019
02/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age62 YR
Patient Weight56
-
-