| Medical Device Problem Code |
Structural Problem (2506)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation the device has been returned and evaluated by product analysis on 25-jul-2019 with the following findings: during investigation, the cartridge compartment was found to be cracked.Unrelated to the original complaint, the battery compartment was found to be cracked.Report late due to transition from vmsr program.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2019, the reporter contacted animas, alleging a casing/condition (cracked/damaged casing) issue.It was reported that the cartridge compartment was cracked/damaged.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the alleged malfunction has the ability to result in a delay in treatment or long-term cessation in delivery if the damage impacts the power circuit or cartridge compartment.
|
| |
|