COOK INC COOK SILICONE BALLOON HYSTEROSALPINGOGRAPHY INJECTION CATHETER; HES INSUFFLATOR-CARBON DIOXIDE, URETOTUBAL (AND ACCESSORIES) Back to Search Results
Model Number G17147
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/08/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported, during an unknown procedure in which a cook silicone balloon hysterosalpingography injection catheter was used, that prior to patient contact, when filling the balloon it ruptured.It is unknown how the procedure was completed.It was reported that the patient did not experience any adverse effects due to this occurrence.Additional information has been requested regarding the patient and the event.At the time of this report, no further information has been provided.
 
Event or Problem Description
No new patient or event information since the last report was submitted on (b)(6) 2019.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unchanged, unknown or unavailable h6: ec method code desc - 5: communication/interview (4111).Investigation - evaluation.Reviews of complaint history, device history record, quality control data and the instructions for use(ifu) were conducted during the investigation.A review of the device history record found no non-conformances related to the reported failure mode.A review of complaint history records shows 1 other complaint associated with the complaint device lot.Due to the individual nature of the manufacturing and inspection process for the devices in the lots, it is unlikely that these events are an indication of device issue within the entire lot.The instructions for use (ifu), provides the following information to the user related to the reported failure mode: precautions- "do not overinflate.Using excessive pressure to inflate the balloon on this device can cause the balloon to rupture." "refer to product label of the inflation check valve on the balloon device for appropriate balloon volume." how supplied- "upon removal from the package, inspect the product to ensure no damage has occurred." the complainant did not return the complaint device to cook for investigation.At the time of the complaint investigation, all products in the complaint device lot have been distributed to customers, so no representative product is available from the lot for evaluation.The complaint was confirmed based on customer testimony.A definitive cause of the event could not be determined.Per the quality engineering risk assessment, no further action is warranted.Cook medical will continue to monitor this device via the complaints database for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK SILICONE BALLOON HYSTEROSALPINGOGRAPHY INJECTION CATHETER
Common Device Name
HES INSUFFLATOR-CARBON DIOXIDE, URETOTUBAL (AND ACCESSORIES)
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key9218430
Report Number1820334-2019-02668
Device Sequence Number16827684
Product Code HES
UDI-Device Identifier00827002171473
UDI-Public(01)00827002171473(17)210906(10)9144491
Combination Product (Y/N)N
Initial Reporter CountryMX
PMA/510(K) Number
K891290
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,other,use
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/06/2021
Device Model NumberG17147
Device Catalogue NumberJ-CHSG-503000
Device Lot Number9144491
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 10/17/2019
Supplement Date Received by Manufacturer11/22/2019
Initial Report FDA Received Date10/22/2019
Supplement Report FDA Received Date11/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-