• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE FB TIB BASE SZ 10 CEM; ATTUNE IMPLANT : KNEE TIBIAL TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE FB TIB BASE SZ 10 CEM; ATTUNE IMPLANT : KNEE TIBIAL TRAY Back to Search Results
Catalog Number 150600010
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 09/30/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The patient was revised to address loosening of the tibial component at the cement to implant interface.Cement manufacturer was unknown.It was also reported that the case was scheduled as a poly exchange vs full revision.When the surgeon went to test the stability of the implants, the tibial tray became loose after a few suttle blows from an osteotome/mallet.The femoral component and the tibial insert were also extracted.The patella component was retained.Doi: (b)(6) 2016; dor: (b)(6) 2019; right knee.
 
Additional Manufacturer Narrative
Depuy synthese is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthese has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthese or its employees that the report constitutes an admission that the device, depuy synthese, or its employees caused or contributed to the potential event described in this report.Product complaint # : (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot : null.Device history batch : null.Device history review : null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE FB TIB BASE SZ 10 CEM
Common Device Name
ATTUNE IMPLANT : KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key9219133
Report Number1818910-2019-111155
Device Sequence Number2223818
Product Code JWH
UDI-Device Identifier10603295042440
UDI-Public10603295042440
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
K101433
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number150600010
Device Lot Number7809757
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/30/2019
Supplement Date Received by Manufacturer11/19/2019
Initial Report FDA Received Date10/22/2019
Supplement Report FDA Received Date11/21/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE PS FB INSRT SZ 10 7MM; ATTUNE PS FEM RT SZ 10 CEM; UNK KNEE PATELLA: UNKNOWN; ATTUNE PS FB INSRT SZ 10 7MM; ATTUNE PS FEM RT SZ 10 CEM; UNK KNEE PATELLA: UNKNOWN
Outcome Attributed to Adverse Event Required Intervention;
-
-