SMITH & NEPHEW, INC. TRL 12/14 TPR FEM HD 32 -3; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
| Catalog Number |
OR71353203 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/20/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that two pieces of inner collar broke off.Occurred inside of patient, all pieces were recovered.S&n backup device was available and was used to complete the surgery.No patient affectation, no pictures available, no closure letter needed, lot number is not on item.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated complaint device was returned to be evaluated.The visual inspection confirms two of connections tabs are broken off the device.The broken pieces were not returned.The device has multiple burrs/gouges/scratches.This device shows signs of extreme wear/usage.The device has no available lot number.We recommend that all re-usable instruments be routinely inspected for wear/damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.
|
| |
|