| Catalog Number |
6125089090 |
| Medical Device Problem Codes |
Break (1069); Fracture (1260)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/24/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A follow up report will be filed once the quality investigation is complete.Awaiting return of device to manufacturer.
|
| |
|
Event or Problem Description
|
|
It was reported that during two separate surgical procedures, it was noted that the blade broke where it fits into the handpiece.It was also reported there were no adverse consequences as a result of this event.It was further reported that a minor delay occurred as a result of the event.It was also reported that the procedure was completed successfully.
|
| |
|
Event or Problem Description
|
|
It was reported that during two separate surgical procedures, it was noted that the blade broke where it fits into the handpiece.It was also reported there were no adverse consequences as a result of this event.It was further reported that a minor delay occurred as a result of the event.It was also reported that the procedure was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results indicate that the blade coming into contact with cut guides/ surgical instruments could have caused the mount tab to fracture at the mount.The quality investigation is complete.
|
| |
|