• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO NAVLOCK, UNIVERSAL TRACKER ADAPTER FOR INSTRUMENTS FROM 7 TO 13MM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO NAVLOCK, UNIVERSAL TRACKER ADAPTER FOR INSTRUMENTS FROM 7 TO 13MM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 6000-999-003
Medical Device Problem Codes Fracture (1260); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 09/27/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during set up it was observed that the washer was broken and missing from the device; this poses the risk of a small component being lost in the surgical site.No medical intervention and no adverse consequences were reported with this event.As this event occurred during setup at the user facility, there was no patient involvement and no delay to a surgical procedure.
 
Additional Manufacturer Narrative
Device evaluation: follow-up report submitted to document device evaluation results.
 
Event or Problem Description
It was reported that during set up it was observed that the washer was broken and missing from the device; this poses the risk of a small component being lost in the surgical site.No medical intervention and no adverse consequences were reported with this event.As this event occurred during setup at the user facility, there was no patient involvement and no delay to a surgical procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVLOCK, UNIVERSAL TRACKER ADAPTER FOR INSTRUMENTS FROM 7 TO 13MM
Common Device Name
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
MDR Report Key9221176
Report Number0001811755-2019-03294
Device Sequence Number16815524
Product Code HAW
UDI-Device Identifier04546540391919
UDI-Public04546540391919
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number6000-999-003
Device Catalogue Number6000-999-003
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 10/02/2019
Supplement Date Received by Manufacturer04/01/2020
Initial Report FDA Received Date10/22/2019
Supplement Report FDA Received Date04/01/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-