BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO (US) 2.25/13; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 401735
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/03/2019
Type of Reportable Event Malfunction
Event or Problem Description
An orsiro drug-eluting stent system was chosen to treat a severely calcified lesion (95 percent stenosis degree) in the tortuous mid lad.After pre-dilatation the orsiro could not pass lesion.An extension catheter was used but the stent still didn't cross.Upon another pre-dilatation and implantation of an 2.75/15 orsiro the complaint instrument was re-inserted but couldn't cross the implanted stent.
 
Additional Manufacturer Narrative
The returned product was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The technical investigation of the returned instrument revealed that the balloon is well folded and shows no signs of inflation.Stent imprints are visible on the balloon surface, indicating that the stent was initially crimped centred on the balloon.The stent was returned separately and is deformed at one end.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.As a part of final inspection every stent system undergoes visual inspection to ensure correct embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent and the stent retention force of a defined amount of samples is tested from every lot.Based on the conducted investigations, no manufacturing or material related root cause could be determined.
 
Event or Problem Description
An orsiro drug-eluting stent system was chosen to treat a severely calcified lesion (95 percent stenosis degree) in the tortuous mid lad.After pre-dilatation the orsiro could not pass lesion.An extension catheter was used but the stent still did not cross.Upon another pre-dilatation and implantation of an 2.75/15 orsiro the complaint instrument was re-inserted but could not cross the implanted stent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO (US) 2.25/13
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key9221665
Report Number1028232-2019-04610
Device Sequence Number2157449
Product Code NIQ
UDI-Device Identifier07640130438903
UDI-Public07640130438903
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/04/2020
Device Model Number401735
Device Catalogue NumberSEE MODEL NO.
Device Lot Number10185745
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2019
Initial Date Received by Manufacturer 10/04/2019
Supplement Date Received by Manufacturer12/17/2019
Initial Report FDA Received Date10/22/2019
Supplement Report FDA Received Date12/17/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age67 YR
-
-