| Model Number |
7580 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/09/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of a maverick2 balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There was contrast in the inflation lumen and in the balloon.Blood was present in the guidewire lumen and balloon.The balloon was loosely folded.The balloon had a pinhole at the distal end of the distal markerband.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed the reported event that the lesion was tight, as there was a pinhole in the balloon.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 07-oct-2019.A 2.00mmx12mm maverick balloon catheter was returned without reported allegation.However, device analysis revealed a balloon pinhole.
|
| |
|