| Model Number |
MMT-723RNAL |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/18/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that insulin pump had motor error alarm past few months when pump runs out of insulin.The blood glucose was unknown.The device will not be returned for the analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed displacement test, rewind test, basic occlusion test, prime/a33 test, excessive no delivery test and occlusion test, no motor error alarm during testing noted.The motor was tested outside the unit and passed.(b)(4).
|
| |
|