Device is a combination product.Device evaluated by mfr.: promus element,mr,ous 4.00x28mm stent delivery system (sds) was returned for analysis.The device was returned inside a 0.071 guide catheter, with a haemostatic valve attached.The stent and balloon were visible distal to the distal tip of the guide catheter.Due to stent damage evident on the device, the device could not be removed proximally from the guide catheter.To facilitate the removal of the device from the guide catheter without further damage to the stent, the investigator cut the hypotube at a location 2mm distal to the distal strain relief.The sds is now in 2 pieces.The device was removed distally without issue.A visual and microscopic examination of the stent found proximal stent damage, with stent struts lifted and bunched proximally an distally.The undamaged stent outer diameter was within maximum crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube found multiple hypotube kinks at several locations along the hypotube shaft.A visual examination of the outer and inner lumen and mid-shaft section found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.
|