| Model Number |
MMT-723RNAS |
| Medical Device Problem Codes |
Mechanical Problem (1384); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/14/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported that they got a replacement cap and after that, they were having issues with motor error.The customer's blood glucose level was unknown at the time of the incident.The customer reported that the drive support cap was protruded.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The device will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received motor error alarm during rewind due to motor encoder signal out of phase.Unable to perform the self-test, a21 error test, displacement test, rewind, basic occlusion test, occlusion test, prime or a33 test and excessive no delivery test due to motor error alarms noted.No loose drive support cap anomaly noted.
|
| |
|