• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER MANUFACTURING B.V. ALL-POLY PATELLA CEMENTED 38 MM DIAMETER; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER MANUFACTURING B.V. ALL-POLY PATELLA CEMENTED 38 MM DIAMETER; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Packaging Problem (3007)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Multiple mdrs were filed for this event.Please see associated: 3007963827 - 2019 - 00303, 3007963827 - 2019 - 00304, 0002648920 - 2019 - 00768, 3007963827 - 2019 - 00302.Concomitant medical products: 42512400714 articular surface fixed bearing posterior stabilized, 64363834.42540200038 all-poly patella cemented 38 mm diameter, 64333036.42540200035 all-poly patella cemented 35 mm diameter, 64435171.42500606401 femur cemented posterior stabilized (ps) standard left size 8, 64295001.42532007101 natural tibia cemented 5 degree stemmed left size e, 6429340.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported during the distribution process packaging damage with sterility barrier potentially compromised was identified.No patient or surgical involvement.No further information available at this time.
 
Additional Manufacturer Narrative
Reported packaging damage event was confirmed via visual examination.It was identified the packaging was damaged in transit.Device history record (dhr) review was performed with no related manufacturing deviations or anomalies identified.Root cause is attributed to transit damage.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further information is available at the time of this reporting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALL-POLY PATELLA CEMENTED 38 MM DIAMETER
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
MDR Report Key9225231
Report Number0002648920-2019-00767
Device Sequence Number1611383
Product Code JWH
UDI-Device Identifier00889024247710
UDI-Public(01)00889024247710
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K172524
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2024
Device Model NumberN/A
Device Catalogue Number42540200038
Device Lot Number64333036
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/25/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 09/25/2019
Supplement Date Received by Manufacturer12/02/2019
Initial Report FDA Received Date10/23/2019
Supplement Report FDA Received Date12/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
-
-