|
On (b)(6) 2019, the mentor failure analysis lab has received the device for evaluation.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: rupture.On (b)(6) 2019, during analysis of the sample was found an area of missed material in the posterior view , measuring approximately 1.4 cm x 1.5 cm and a tear in the anterior view, measuring approximately 4.5 cm.Microscopic examination was performed in both tears, and no indications of instrument damage was found.No other anomalies were discovered.The manufacturing records were reviewed and the manufacturing criteria were met prior to the release of this lot.Rupture, as indicated in the product insert data sheet (pids), is a known complication associated with mentor mammary prostheses.Causes of rupture of gel-filled implants include, but are not limited to the following events: damage from surgical instruments, intraoperative or postoperative trauma,excessive stresses or manipulations as may occur during normal, daily routines including customary and purposeful trauma to the breast.Rupture is a known complication associated with these devices and is referenced in our product insert data sheet.It should be noted that as part of our quality process, each device is visually inspected during manufacturing to ensure the device meets the required specifications prior to shipment.Based on the information reported and/or the product investigation, there is no evidence that the issue is related with the manufacturing process.Complaint information will be included in complaint trending that is reviewed and monitor by quality assurance on a regular basis to determine if further action is necessary.Manufacturer¿s reference number: (b)(4).
|