• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NIPRO (THAILAND) CORP. LTD. NIPRO BLOOD TUBING SET W/TRANSDUCER PROTECTOR; BLOODLINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIPRO (THAILAND) CORP. LTD. NIPRO BLOOD TUBING SET W/TRANSDUCER PROTECTOR; BLOODLINE Back to Search Results
Model Number BL+A209Y/V803
Health Effect - Clinical Code Blood Loss (2597)
Date of Event 09/25/2019
Type of Reportable Event Malfunction
Event or Problem Description
0740am patient care technician noticed blood leak on the blood pump area of the machine.Patient care tech notified the facility administrator, upon inspection noted the tubing on the blood pump had a tear and was leaking blood, staff rinsed back the arterial line, was unable to rinse back venous line due to the leak, patient lost <150cc of blood, staff explained to the patient what had happened, and advised the patient to report any signs and symptoms of anemia, will redraw labs next treatment, new blood lines were set up, and treatment was resumed at 0747am.Blood pressure was within limits, patient denies any complaints.Physician was made aware.Lot number was unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIPRO BLOOD TUBING SET W/TRANSDUCER PROTECTOR
Common Device Name
BLOODLINE
Manufacturer (Section D)
NIPRO (THAILAND) CORP. LTD.
10/2 moo 8, bangnomko, sena
ayuthaya,, 13110
TH  13110
MDR Report Key9226896
Report Number1056186-2019-00012
Device Sequence Number2549545
Product Code FJK
UDI-Device Identifier00383790004053
UDI-Public00383790004053
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/23/2019
Report Date (Section F) 10/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBL+A209Y/V803
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date09/28/2019
Event Location Outpatient Treatment Facility
Date Report to Manufacturer10/23/2019
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date10/23/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Date Report Sent to FDA10/23/2019
-
-