| Catalog Number |
381467 |
| Medical Device Problem Codes |
Leak/Splash (1354); Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
09/19/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medical device expiration date: unknown.A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.Device manufacture date: unknown.
|
| |
|
Event or Problem Description
|
|
It was reported that bd insyte¿ autoguard¿ shielded iv catheter separated from the hub during use.The following information was provided by the initial reporter: material no.: 381467 batch no.: unknown.It was reported that the catheter separated from the hub of the iv and remained in the patient's arm.Good morning, there was an event reported on rhoads 4 that needs f/u.We do not have the catheter.I have copied clinical leadership on this email in order to expedite the information gathering process.Please keep in mind we do not share patient information.Patient had a left ac 14g iv.Iv no longer indicated.Upon removing iv, catheter separated from hub of iv and remained in patient's vein.Transplant md came to bedside and was able to retrieve catheter.Patient required small incision to remove catheter that was then covered with a steristrip.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: the provided photo shows a yellow adapter unit which had an extension line attached to the end of the adapter (luer).There were traces of thick media present.The yellow unit in the photo was not acknowledged as the actual sample because the reported part number (381467) noted pertained to a 14ga unit which is of the color orange.The appearance of the reported device was compared to a catheter adapter of the same product based on the reported part number and it did not match: bd autoguard 381467 is yellow, has a taper, a tab and grips.The reported adapter does not have these features.In fact, the reported catheter adapter does not look like any of the bd sandy products.Conclusion(s): not determined ¿ the investigation concluded that the complaint is not related to bd autoguard product.
|
| |
|
Event or Problem Description
|
|
It was reported that bd insyte¿ autoguard¿ shielded iv catheter separated from the hub during use.The following information was provided by the initial reporter: material no.: 381467 batch no.: unknown.It was reported that the catheter separated from the hub of the iv and remained in the patient's arm.Good morning, there was an event reported on rhoads 4 that needs f/u.We do not have the catheter.I have copied clinical leadership on this email in order to expedite the information gathering process.Please keep in mind we do not share patient information.Patient had a left ac 14g iv.Iv no longer indicated.Upon removing iv, catheter separated from hub of iv and remained in patients vein.Transplant md came to bedside and was able to retrieve catheter.Patient required small incision to remove catheter that was then covered with a steristrip.
|
| |
|