Catalog Number 1012624-39
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/30/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Exemption number e2019001.The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the procedure was to treat a moderately calcified lesion in the common iliac artery with no tortuosity.No resistance was noted during advancement and the balloon of the 8x39 mm omnilink stent delivery system (sds) was inflated twice and ruptured at nominal pressure.The stent was successfully implanted.There was no issue removing the balloon and there were no adverse patient effects.There was no reported clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
Exemption number e2019001-permits numbering sequence to begin with 10000, to avoid duplication of report numbers due to process transition.There may be gaps in numbering for reports submitted during the transition period.Visual analysis was performed on the return device.The reported balloon rupture was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined that the reported balloon rupture was due to case circumstances.It is likely that the balloon rupture was the result of interaction with the calcified lesion.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNILINK ELITE
Common Device Name
PERIPHERAL STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key9227443
Report Number2024168-2019-12901
Device Sequence Number1564734
Product Code NIO
UDI-Device Identifier08717648179303
UDI-Public08717648179303
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
P110043
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2022
Device Catalogue Number1012624-39
Device Lot Number9010441
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/29/2019
Initial Date Received by Manufacturer 10/01/2019
Supplement Date Received by Manufacturer11/16/2019
Initial Report FDA Received Date10/23/2019
Supplement Report FDA Received Date12/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
6F SHEATH
-
-